Cadmium content testing methods are divided into two categories: total cadmium testing and cadmium migration testing.
I. Total Cadmium Content Test
Reference StandardEN 1122:2001 (Determination of Cadmium in Plastics - Wet Digestion), IEC 62321-5 (Determination of Cadmium in Electrical and Electronic Products - Acid Digestion), CPSC-CH-E1001-08.3 (Metal Parts), CPSC-CH-E1002-08.3 (Non-metal Parts)
Testing ProcessAfter the sample is crushed, microwave digestion is performed using concentrated sulfuric acid + hydrogen peroxide (EN 1122 plastics method) or concentrated nitric acid + hydrogen peroxide (IEC 62321 general method). After dilution, the digested solution is quantitatively analyzed using ICP-OES, ICP-MS, or AAS. The detection limit is typically 2 to5 ppm。
II. Cadmium Migration Test
Reference StandardEN 71-3 (Toys, extraction with 0.07 mol/L hydrochloric acid simulating gastric acid), 84/500/EEC (Ceramics,4 %soaked in acetic acid for 24 hours).
Testing ProcessThe sample is immersed in the extraction solution, and after extraction under specified conditions, the cadmium content in the extraction solution is determined using ICP-OES or ICP-MS.
III. XRF Rapid Screening
XRF can be used to rapidly determine the approximate content of cadmium, which is suitable for large-batch screening. However, accurate analysis still requires chemical digestion methods.